Chapter 12: IBP Monitoring
User Manual
12-9
■
IBP ZERO
Pick < ZERO> in IBP WINDOW to call up the following menu:
NOTE:
Zero procedure should be performed before starting the monitoring and at least
once a day after each disconnection and connection of the cable.
Zero the transducer:
1-The transducer should be placed at mid-heart level.
2-Turn off patient stopcock.
3-The transducer must be vented to atmospheric pressure.
4-Select < IBP1/IBP2 ZERO > to start zeroing procedure for each channel.
The message "PLEASE WAIT" will be displayed during the procedure. When the procedure
finished successfully the message "IBP1/IBP2 ZERO OK" appears.
The last zeroing time will be saved and displayed in its corresponding place.
5-Turn stopcock to patient on and the other stopcock to atmospheric pressure off.
The following messages may prompt up in ZERO WINDOW:
■
"IBP1/IBP2 NO SENSOR, UNABLE TO ZERO"
Make sure that the transducer is connected or not, then start zeroing.
■
"IBP1/IBP2 OVERANGE, FAILED ZEROING"
Make sure that the stopcock is vented to atmosphere. If the problem persists, contact
After Sale Service.
■
"IBP1/IBP2 UNSTABLE PRESSURE, UNABLE TO ZERO"