Chapter 22:Technical Specifications
User Manual
22-9
Manual compensation for broadening effects on CO2.
ISA OR+/AX+
Automatic compensation for pressure, temperature and
broadening effects on CO2.
Warm-up time
ISA CO2:
<10 seconds (concentrations reported and full accuracy)
ISA OR+/AX
+: <20 seconds (concentrations reported, automatic agent
identification enabled and full accuracy
Rise time
9
CO2
200 ms(
300 ms for ISA
OR+/AX+)
N2O, O2, ENF, ISO, SEV, DES
400 ms
HAL
500 ms
Primary agent threshold (ISA OR+/AX+)
0.15 vol%. When an agent is identified, concentrations will be reported even
below 0.15 vol%
Secondary agent threshold (ISA
OR+/AX+)
0.2 vol% +10% of total agent concentration
Agent identification time (ISA OR+/AX+)
<20 seconds (typically <10 seconds)
Total system response time
1
0
ISA CO2:
< 3 seconds
ISA OR+/AX+:
< 4 seconds (with 2m Nemoline Airway Adapter Set
sampling line)
Accuracy standard conditions
The following accuracy specifications are valid with no drift for dry single gases at 22 ± 5 °C and 1013 ± 40 hPa:
Range
1
1
Accuracy
CO2
0 to15 vol%
±(0.2 vol% +2% of reading)
N2O
0 to 100 vol%
±(2 vol% +2% of reading)
HAL, ENF, ISO
0 to 8 vol%
±(0.15 vol%+5% of reading)
SEV
0 to 10 vol%
±(0.15 vol% +5% of reading)
DES
0 to 22 vol%
±(0.15 vol% +5% of reading)
O2
0 to 100 vol%
±(1 vol% +2% of reading)
Accuracy - all conditions
The following accuracy specifications are valid with no drift for all specified environmental conditions, except for
interference from water vapor in the below section “Effects from water vapor partial pressure on gas readings”.
GAS
Accuracy
CO2
±(0.3 kPa + 4% of reading)
N2O
±(2 kPa + 5% of reading)
Agents
1
2
±(0.2 kPa + 10% of reading)
O2
±(2 kPa + 2% of reading)
Effects from water vapor partial pressure on gas readings
When the breathing gas flows through the sampling line, the gas temperature will adapt to the ambient temperature before
reaching the gas analyzer. The measurement of all gases will always show the actual partial pressure at the current humidity
level in the gas sample. As the NOMO section removes all condensed water, no water will reach the ISA gas analyzer.
However at an ambient temperature of 37 °C and a breathing gas with a relative humidity of 95% the gas reading will
typically be 6% lower than corresponding partial pressure after removal of all water.
Interfering gas and vapor effects
9
Measured according to EN ISO 80601-2-55.
1
0
Measured according to EN ISO 80601-2-55.
1
1
All gas concentrations are reported in units of volume percent and may be translated into mmHg or kPa by using the reported
atmospheric pressure.
1
2
The accuracy specification is not valid if more than two agents are present in the gas mixture. If more than two agents
are present, an alarm will be set.