Chapter 13: GAS Monitoring (Mainstream)
User Manual
13-16
EtCO2 LIMIT
Alarm is activated when the EtCo2 exceeds adjusted ALARM HIGH limit or falls below
adjusted ALARM LOW limit. (Range: 0.4~13%V step 0.1%V)
Default for upper limit is 6.5%V and for lower limit is 2.6%V.
FiCO2 HIGH
Alarm is activated when the FiCo2 exceeds adjusted ALARM HIGH limit.
(Range: 0.4~13 %V step 0.1%V), Default for upper limit is 1.3%V.
AWRR LIMIT
Alarm is activated when the AWRR exceeds adjusted ALARM HIGH limit or
falls below adjusted ALARM LOW limit.(Range :1-120BrPM )
Default for upper limit:
Adult/Pediatric:
30BrPM
Neonate:
60BrPM
Default for lower limit:
Adult/ Pediatric:
5BrPM
Neonate:
15BrPM
APNEA LIMIT
Pick it to set the standard of judging an apnea case. It sets to 20 - 60 seconds and "OFF",
increases/decreases by 5s. Select OFF to disable alarm.
■
GAS ALARM
Pick "GAS ALARM" in GAS WINDOW to call up the following menu:
GAS/ALARM WINDOW