Chapter 11: TEMP Monitoring
User Manual
11-4
The TEMP WINDOW is as follows:
■
UNIT
Pick this item to set measurement unit. (options: ºC or ºF)
■
ALARM LEVEL
Selectable between 1 and 2. Level 1 represents the most serious case.
■
TEMP ALARM
Pick "ON" to enable TEMP alarm functions such as parameters blinking, audio alarm, and
light indicator.Pick "OFF" to disable the alarm functions and there will be a "
" symbol
in the Parameter Area.
■
T1 LIMIT
T1 alarm is activated when the channel-1 temperature exceeds adjusted ALARM HIGH limit
or falls below adjusted ALARM LOW limit.(min:0.0 and max:50.0)
■
T2 LIMIT
T2 alarm is activated when the channel-2 temperature exceeds adjusted ALARM HIGH
limit or falls below adjusted ALARM LOW limit.(min:0.0 and max:50.0)
■
DT LIMIT
DT alarm is activated when the difference between channel-1 and channel-2 exceeds adjusted
ALARM HIGH limit or falls below adjusted ALARM LOW limit. (min:0.0 and max:50.0).
T1 is Channel-1 of temperature.
T2 is Channel-2 of temperature
DT is the temperature difference between the above two.