Chapter 22:Technical Specifications
User Manual
22-2
CLASSIFICATION
Protection against electroshock
Class I, Type CF for all modules (except Multi-gas, NIBP and BFA
modules that are BF) (based on IEC 60601-1).
Protection
Against Electro surgery and Defibrillator (Except BFA)
Mode of operation
Continues operation equipment
Harmful Liquid Proof Degree
IPX1
Method of disinfection
Refer to each module's chapters and chapter Care & Cleaning for detail.
Safety of anesthetic mixture
Not suitable for use in the presence of a flammable anaesthetic mixture
with air or with oxygen or nitrous oxide.
General
Display
Alborz B9
TFT/LED
COLOR
1366
768
18.5”
Waveforms
ECG, SpO2, IBP1, IBP2, RESP/GAS,EEG (Freezable), C.O.
Numeric Parameters
HR,PVCs,ST,SpO2, PR, NIBP (SYS, DIA, MAP),
IBP1(SYS,DIA,MAP), IBP2(SYS,DIA,MAP), RR, T1, T2, DT, EtCo2,
FiCo2, AWRR, EtN2O, FiN2O, EtO2, FiO2, EtAA, FiAA, BFI, BS%,
EMG%, SQI%, C.O, Alarm Limits.
Operation Method
Membrane/Keys and rotary knob, Touch Screen
AC Power
100 - 240 VAC, 50/60 Hz , Ip: 0.9
–
0.4 A
ECG
Lead & Wire Options
Selectable: 3 ,5 or 12 Leads
Selectable: 3 ,5 or 10 Wires
3 ECG Leads I, II, III
3 Lead wires ECG Cable
5 Leads ECG: I,II,III,V,aVR,aVF,aVL
5 Lead wires ECG Cable
12 Leads ECG : I,II,III, aVR,aVF,aVL , V1,V2, V3, V4, V5. V6
10 Lead wires ECG Cable
Dynamic Range
5 mV
Lead Off Current
< 90 nA
Gain
4, 2, 1, 1/2, 1/4, Auto
Calibration
1mV, 0.5 sec
Filters
“MONITOR”
0.5 - 24 Hz
“NORMAL”
0.5 - 40 Hz
“EXTENDED”
0.05-100 Hz
CMRR
> 98 dB
Internal Noise
< 30
µ
V RTI
Input Impedance
> 5 M
Ω
QRS Detection
Duration
40 to 120 msec
Amplitude
0.25 to 5 mV
for Adult/Pediatric
0.2 to 5 mV
for Neonate
Heart Rate Range
15 - 300 BPM
for adult/Pediatric
15 - 350 BPM
for neonate
Accuracy
1% or 2 BPM
Tall T-Wave
Reject up to 1.2 mV Amp.
Pacer Detection/Rejection
Duration
0.1 - 2 msec
Amp
±
2 to
±
700 mV (Without over/undershoot)