Chapter 16: Cardiac Output Monitoring
User Manual
16-3
NOTE:
Fluid injection must be carried out smoothly.
Warning
C.O. measurement must be carried out by trained and qualified individuals.
C.O. value is influenced by injection technique. As mentioned above, the curve should have a
steep rise and gradual return to baseline (See figure 15-2). Other curves are samples of high
cardiac output, low cardiac output and prolonged injection. A series of measurements must be
carried out to achieve a reliable C.O. value and average of multiple thermodilution
measurements is used for therapy decisions.