Chapter 18: Recording
User Manual
18-9
18.6 Recorder Alarm Messages
Message
Cause
Solution
Rec. Software Error
Software error
Turn the system off and then on .If the problem
persists, contact after sales service of manufacturer.
Recorder Fault
Hardware error
Turn the system off and then on. If the problem
persists, contact after sales service of manufacturer.
Rec Door Open
The recorder door is open
Close the recorder door.
Rec Paper Out
Recorder paper has been finished.
Insert a new paper roll.
Print head High Temp
The thermal head is too hot.
Stop operation for some minutes.
Print head High Vol.
Print head voltage is high.
Turn the system off and then on. If the problem
persists, contact after sales service of manufacturer.
Print head Low Vol.
Print head voltage is low.
Turn the system off and then on. If the problem
persists, contact after sales service of manufacturer.
Time out Error
The recorder could not record.
Turn the system off and then on. If the problem
persists, contact after sales service of manufacturer.
All alarm messages of the recorder are level 2. The messages are displayed in the yellow
background. By pressing Silence key, the message background will change to the gray and the
system will ignore this fault.