Chapter 5: ECG Monitoring
User Manual
5-23
5.6 ECG Alarms
a) Physiological alarms
The auditory alarm sounds when:
1. The heart rate violates the adjusted alarm limits, and/or,
2. The ECG ASYSTOLE occurs.
ALARM
SITUATION
VISUAL PROMPTS
AUDIO SOUND
HR HIGH
Heart rate violates
adjusted high limit
●HR value blink
s.
●The alarm indicator flashes.
●
The alarm message is
displayed in a background
corresponding to its level.
Activated
HR LOW
Heart rate violates
adjusted low limit
●HR value blinks.
● The alarm indicator
flashes.
●
The alarm message is
displayed in a background
corresponding to its level.
Activated
ECG ASYSTOLE
Heart beat is not detected
in last 10 seconds.
●HR is "00" and blinks
● The alarm indicator
flashes.
● The message "ECG
ASYSTOLE" is displayed
in red background.
Activated