Chapter 14: GAS Monitoring (Sidestream)
User Manual
14-18
FiN2O LIMIT
Alarm is activated when the FiN2O exceeds adjusted ALARM HIGH or LOW limit.
(Range: 1~82%V, step1%V)
Default for upper limit is 75%V and for lower limit is 35%V.
EtAA LIMIT
Alarm is activated when the EtAA exceeds adjusted ALARM HIGH or LOW limit.
FiAA LIMIT
Alarm is activated when the FiAA exceeds adjusted ALARM HIGH or LOW limit.
NOTE:
Alarm range and alarm limit default of different anesthesia agents are
mentioned in the table below:
Anesthesia agent
Alarm range
Step
Alarm limit default
HAL
0.1~5%
0.1%
0.5~1.5%
DES
0.1~18%
0.1%
5~10%
ISO
0.1~5%
0.1%
0.8~2%
SEV
0.1~8%
0.1%
1~3%
ENF
0.1~5%
0.1%
0.5~1.5%
EtO2 LIMIT
Alarm is activated when the EtO2 exceeds adjusted ALARM HIGH or LOW limit.
(Range: 18~105%V step1%V)
Default for upper limit is 100% and for lower limit is 50%.
FiO2 LIMIT
Alarm is activated when the FiO2 exceeds adjusted ALARM HIGH or LOW limit.
(Range: 18~105%V step 1%V)
Default for upper limit is 100%V and for lower limit is 50%V.