Chapter 15: BFA Monitoring
User Manual
15-11
■
To enable or disable the BFA alarm limit:
Pick
“
BFA
ALM ON/OFF” to enable
BFI alarm function such as parameters blinking, audio
alarm and light indicator.
Pick “OFF” to disable the alarm functions and there will be a
will be a
"
" symbol in the Parameter Area.
■
To set the BFI alarm limit:
Press the 10
th
left item to set the BFI limit. Alarm is activated when the BFI parameter exceeds
adjusted ALARM HIGH limit or falls below adjusted ALARM LOW limit. (default: min= 35,
max=60)
NOTE:
BFI alarm level is always II.
NOTE:
Every change in BFA large page setting is seen in BFA window in normal state.
NOTE:
In case of sudden and strange changes in BFI or SQI index, it is necessary to
measure impedance manually.