O-ring replacement
1. Use the needle tool TLAX110023 to loosen the outer needle from the needle holder.
Warning:
Be aware of spring tension
2. Store outer needle in tool TLAX110023.
3. Remove the spring QWVR0012260.
4. Loosen TLAX090012 O-ringscrew from the needle holder.
5. Remove the O-ring QWLV050002.
6. If needed: Grease the outside of the inner needle base by removing TLAX090011 (use acid free vaseline).
7. Fit a new O-ring QWLV050002 carefully, and press the O-ring into the recess.