1. Pull pump tube slightly downwards and at the same time towards the front of the unit to release the tube out
of the pump plate holder.
2. Remove the old tube from the peristaltic pump rotor.
3. Disconnect the tubing at both ends of the tube connectors.
4. Connect new tubing to both ends of the connectors.
5. Place one end of the tube in the pump plate holder.
6. Pull the new tube over the peristaltic pump rotor.
7. Pull pump tube slightly downwards and at the same time towards the back of the Interrliner.
If the tube is not fitted correctly or is worn the following symptoms can occur:
▪
Liquid flowing back into the container.
▪
First glass tube on the pipette belt is not washed sufficiently.
Note:
The wider bore tube is for the pump A.
Waste pump tube QWLV090011 in cassette.
▪
Remove the old tube from the pump as shown:
▪
Pull the tube from the connectors and install the new tube in the reverse order.
▪
Check that all connecting tubes and the tube cover are attached.
Work instruction Interrliner
MRN-164_2-EN Version 1.08 Starrsed TL
107
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...