1. Remove the silicone tube (circled in red) from the middle slot of the Liquid sensor. Check the outside of the
tube and check each slot closely to ensure no white substance or dust is blocking the LEDs. If cleaning is
needed, continue with step 2.
2. Clean the tube near the Liquid sensor with De-ionized water and dry with canned air or a gauze pad.
3. Remove the drip tray under the needle.
4. Use a soft fine bristle toothbrush to remove any foreign particles in front of the sensor LEDs. Brush up and
down in the slots on the left and right side of the tube space (shown in red circles).
5. Replace the silicone tube into middle slot of Liquid sensor and test under
Maintenance - Check sensors-Check
Liquid Sensor-Test.
The
Rinse
and
Drain
should show two blue check marks.
Work instruction Interrliner
112
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...