10.2.3 - Cleaning liquid separator
The separator is designed to separate liquid from the air and can handle a lot of blood, rinse and other used
reagents from the instrument. After a period of time the separator is getting dirty and therefore it needs to
be cleaned weekly.
Detailed instructions of this procedure can be found in the Work Instruction
.
Symptoms of a dirty separator:
1. Separator errors.
2. Foam in the separator.
3. Waste pump cannot sufficiently remove the waste from the separator.
10.3 - Level 4 maintenance
The purpose of level 4 maintenance is to carry out the daily / weekly maintenance and replace the pump
tubing, bacterial filters and the Fill nozzle O-ring. After replacing those items, the instrument needs a Fill and
Clean sequence to clean the pipettes. Over a period protein builds up in the Westergren pipettes and needs
to be deproteinized using a strong cleaning agent.
Also the Liquid sensor have to be checked or cleaned of needed.
Detailed instructions of this procedure can be found in the Work Instruction
.
10.3.1 - Replace bacterial filters
Detailed instructions of this procedure can be found in
As part of the Cleaning liquid separator procedure the bacterial Hepa filter
QWLV040002
is replaced with a
new one.
10.3.2 - Fill-nozzle O-ring replacement
As the fill nozzle O-ring (
QWLV050004
) ages, it looses its flexibility and air-bubbles may occur in the
Westergren pipettes, the washer needs to be replaced.
Symptoms for a bad fill-nozzle O-ring
After the aspiration, the Westergren pipette has a zebra pattern (air- blood- air -blood, nicely divided in the
column.)
Vacuum stabilisation errors may occur.
10.3.3 - Fill and clean procedure
Note
: Each pipette on the pipette belt will be filled with Starrsed Cleaning Agent, after one hour the first
pipette is washed and dried. Fill and clean takes about 1 ½ hours to complete.
Fill and clean with adapter:
1. Fill beaker with hot de-ionized water (± 130 ml, ±80°C).
2. Add ±15 ml cleaning agent (QRR 010905) to the hot de-ionized water.
3. Mix the solution well.
4. Fill adapter TLAX110906 with the cleaning solution.
- Turn adapter upside down and remove red cap (from dummy tube) and start filling the adapter until the
dummy sample tube is filled without air.
Maintenance
68
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...