C. A little silicon grease on the rim of separator will make the assembling and adjustment easier.
D. If applicable replace the bacterial HEPA filter (For Maintenance Level 4: Exchange filter QWLV040002)
Replacing
Exchange peristaltic pump tubes
Fill and clean
1. Fill beaker with hot de-ionized water (± 130 ml, ±80°C).
2. Add ±15 ml cleaning agent (QRR 010905) to the hot de-ionized water.
3. Mix the solution well.
4. Fill adapter TLAX110906 with the cleaning solution.
- Turn adapter upside down and remove red cap (from dummy tube) and start filling the adapter until the
dummy sample tube is filled without air.
- Place red cap on dummy sample tube and turn adapter again, fill the adapter completely. Check for air in the
dummy sample tube.
5. Ensure that the Interrliner is in Service mode, the status is also indicated with orange LED light.
6. Enable "Sample Mode On"
Work instruction Interrliner
98
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...