Check sensors
Vacuum pressure check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Check Flow sensor
box.
Flow:
0980
± 60 Abs:
0320
±10
If the flow is not in range there might be a blockage in the vacuum flow line to the flow sensor.
Fill Stop sensor check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Check Fill stop sensor
box.
Fill stop sensor FS 90..
140
..165
Diluter Start sensor check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Diluter start sensor
box.
Diluter start sensor 400-700
Measure sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check measure sensor
box.
Measure sensor MS .
50
±10
Temperature sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Temperature sensor
box.
Temperature sensor TS [Room temperature]
Diluent flow sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Diluent flow sensor
box.
Press test. When test is finished, signal Standby and Flow must shown as activated.
Separator check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Separator sensor
box.
Separator sensor <200
600
>700
Liquid sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check liquid sensor
box.
Check liquid sensor: Press test. When test is finished, the OK signs are shown at Rinse and Drain if the liquid sensor is active.
Check and clean instrument
Prepare disinfectant:
(if not already prepared).
This disinfectant is for cleaning of all external parts that are exposed to blood.
1. Go to tab
[Maintenance]
and perform the
[
End-of-day wash
]
procedure.
2. Open cover and front doors.
3. Check system for leakage.
4. Check drip trays under pump assembly and needle assembly .
5. Open carousel plate and inspect the peristaltic pump tubes and connections for leaks.
6. Check tubing from the syringe for trapped air bubbles.
7. Check Diluent syringe for trapped air bubbles.
8. If trapped air bubbles are found, go to tab
[Maintenance],
click button
[Prime / Clean]
and perform the
[Prime
Diluent]
function.
Work instruction Interrliner
94
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...