E13: Fillnozzle not in home position!
Fill nozzle did not reach the Home
position within a certain time limit.
•
Fill nozzle motor is faulty.
•
Fill nozzle motor driver on
nozzle board is faulty.
•
Fill nozzle is blocked.
E14: Needle motor position error!
(home)
Needle did not reach the home (top)
sensor within a certain time limit.
•
Check home sensor.
•
Needle motor is faulty.
•
Needle motor driver on
needle board is faulty.
•
Needle is blocked.
E18: Carousel position error! Check
Rinse position.
Value of potentiometer does not match
the value stored in memory of the
current rinse position.
•
Check if the rinse position is
right.
•
Set correct rinse position and
perform a "".
•
Check mechanical connection
potentiometers.
E19: Drive motor timeout!
Drive motor did not move or motor is
blocked
•
Check the home sensor
•
Motor is faulty
•
Motor driver on drive board is
faulty
E29: Result path not found. Switched
to default (D:\). Check "Result Path"
setting.
Selected result path is not valid.
Software is using the default setting
•
Check result path setting
•
Check if network or USB
devices are used.
E30: No ACK/NACK received from host
after sending inquiry!
No response from Host within a certain
time limit after sending an inquiry 3
times.
•
Check communication cable
between Host and Interrliner
computer.
•
Check serial port settings
(baud rate, etc...)
•
Check protocol settings.
•
Check Host computer.
E31: NACK received from host after
sending inquiry!
Did not receive ACK from Host after
sending inquiry 3 times.
See E30
E32: LIMS Connection timeout. Host
not found!
The ESR analyser unit could not
establish a connection with the HOST
(server) via TCP/IP.
•
Check TCP/IP settings
•
Check network cable
•
Check HOST settings
E34: No response from host after
sending 'Sample data record'!
No response from Host within a certain
time limit after 3 attempts.
See E30
E35: No response from host after
sending 'Sample flag record'!
No response from Host within a certain
time limit after 3 attempts.
See E30
Appendix for Interrliner
MRN-164_2-EN Version 1.08 Starrsed TL
127
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...