11.3.1 - Reagents alarm
The software checks the reagent level status before starting a new aspiration. If a level alarm is
ON
, it will
not process the new sample. If an alarm comes
ON
during an aspiration, it will finish to aspirate that
sample. Washing dirty pipettes always continues, as to avoid that the samples are left in the pipettes.
Reagents alarm is also set when the expire date of the reagent is exceeded or opened more than three
months. The message Not allowed now! See
Reagents
! appears. Processing of new samples is stopped.
11.4 - Separator error
If it takes too long for the waste pump to empty the liquid separator, the system generates a separator error.
Separator error may be caused by:
Extensive foam build-up in the liquid separator.
Check the separator assembly and connections for possible air
leaks.
Waste-tube between liquid separator and waste
pump is blocked.
Replace the tube.
Waste pump failure.
Check waste pump. If the error returns, call for service.
Waste line blocked.
Check waste line.
Electrical bridge between the waste-level
electrodes.
Clean liquid separator, see
11.5 - Fill time-out error
Normally the fill sequence takes about 3 seconds. However, if the fill sequence exceeds 10 seconds, a fill
time-out error will be generated. The ESR analyser unit aborts the fill sequence and this error message will
be shown on the display and reported.
Fill time-out error may be caused by:
Troubleshooting
78
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...