Check Flow sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
▪
Check liquid sensor: Press test. When test is finished, the OK signs are shown at Rinse and Drain if the
liquid sensor is active.
▪
Check Measure sensor:
Clean sensor first before executing this function.
Click the Check button, the OK sign is shown if the sensor value is in range.
Press the button
Measure
. The pipette currently at the measure position will be measured.
The results are displayed in graphical form:
Measure head start
position correct
Measure head start
position wrong
NOTE:
When a test pipette is installed at the measuring position the result of the test pipette is displayed in the
field "ESR (mm)".
When the sensor is out of range and a red "X" is shown, the sensor values can be
checked by turning on the service mode.
Starrsed TL program
50
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...