Clean and check instrument
1. Check system for leakage.
2. Check drip trays under pump assembly and needle assembly.
3. Open carousel plate and inspect the peristaltic pump tubes and connections for leaks.
4. Check tubing from the syringe for trapped air bubbles.
5. Check Diluent syringe for trapped air bubbles.
6. If trapped air bubbles are found, go to tab
[Maintenance],
click button
[Prime / Clean]
and perform the
[Prime
Diluent]
function.
7. Clean needle top with disinfectant.
8. Wipe stainless steel plates below the pipettes with disinfectant.
9. Close cover/front doors and activate the instrument in the LAS.
Air filter replacement
Air filter replacement QWLV040003
1. Pull both tube connectors out of the blue filter.
2. Place new blue filter
3. Reconnect the tube connectors on the filter
Work instruction Interrliner
114
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...