9. Clean needle top with disinfectant.
10. Check if the suction tube in the guide fork is free of clogging, remove clogging if needed.
11. Wipe stainless steel plates below the pipettes with disinfectant.
12. Close cover/front doors and activate the instrument in the LAS.
Work instruction Interrliner
MRN-164_2-EN Version 1.08 Starrsed TL
95