E270: Needle system failure. No liquid
detected in needle tube during rinse
cycle!
No rinse liquid flow detected in the last
30 cycles. Sample mode stopped.
•
Check Liquid sensor
•
Check tubing
E271: Needle system failure. Liquid
detected in needle tube after draining!
Drain cycle not completed during
drying. Sample mode stops.
•
Check Liquid sensor
•
Check tubing
E272: Needle system failure. Liquid
detected in needle tube before
sampling!
Drain failures detected during drying.
Sample mode stops.
•
Check Liquid sensor
•
Check tubing
Appendix for Interrliner
MRN-164_2-EN Version 1.08 Starrsed TL
133