E258: Error while placing a tube!
Something went wrong while placing a
tube. Clear this error to see the
specific error prior to this error.
•
Clear error to see specific
error.
E259: Gripper tube detect sensor or
gripper tube end sensor already
triggered!
Not possible to pick up a tube, the
gripper tube detect sensor or the
gripper tube end sensor was already
triggered.
•
Faulty gripper tube hat
adjustment.
•
Faulty gripper sensor(s).
E260: After picking up a tube, gripper
tube end sensor still triggered!
After picking up a tube, the gripper
tube end sensor is still triggered.
•
Faulty gripper tube hat
adjustment.
•
Faulty gripper end sensor.
E261: Tube is still present at the
concerning location (cap could be
removed from tube)!
After picking up a tube, the tube is still
present at the concerning location.
Warning: the cap could be removed
from the tube!
•
Location holds the tube too
strong.
•
Incorrect tube (e.g. diameter
too big).
•
Faulty position calibration
(incorrect depth).
•
Faulty tube detect sensor at
the concerning location.
E262: Tube not present in gripper !
After trying to pick up a tube, the tube
is not present in the closed gripper.
Tube maybe lost or still at the
concerning location.
•
Faulty gripper sensor(s).
•
Faulty internal mechanics of
the gripper.
•
Faulty position calibration
(incorrect depth).
•
Tube lost.
E263: Tube too high in gripper before
placing!
The gripper tube end sensor was
triggered too soon while placing a
tube. Another tube may already be
present at the concerning location
(tube on tube).
•
Tube stuck in gripper.
•
Tube on tube.
•
Faulty gripper tube hat
adjustment.
•
Faulty gripper sensor(s).
E264: Not possible to place a tube,
already another tube present!
Not possible to place a tube at the
concerning location, another tube is
already present at that location.
•
Remove tube at concerning
location.
•
Faulty tube detect sensor at
the concerning location.
E265: Still tube detected in gripper
after placing!
After placing a tube, the gripper tube
detect sensor or the gripper tube end
sensor is still triggered. Tube may still
be in the gripper.
•
Faulty gripper sensor(s).
•
Faulty internal mechanics of
the gripper.
•
Tube stuck in gripper.
E266: Tube not present at location
after placing!
After placing a tube, the tube is not
present at the concerning location.
Tube maybe lost or still in the gripper.
•
Faulty tube detect sensor at
the concerning location
•
Tube lost.
•
Tube stuck in gripper.
Appendix for Interrliner
132
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...