Troubleshooting
R&S
®
NRX
493
User Manual 1178.5566.02 ─ 08
Possible reasons:
●
Insufficient airflow. Follow the instructions in
"To place the R&S NRX on a bench
on page 18 or
Chapter 3.1.5, "Mounting in a Rack"
●
The environmental temperature exceeds the suitable temperature range given in
the data sheet under environmental conditions.
●
The fan does not work properly. See
Chapter 16.5, "Contacting Customer Support"
1009 - fan failure alert
The fan does not work. Overheating can damage the R&S NRX.
Solution: Switch off the R&S NRX, and contact customer support. See
16.2.1.2
Power Sensor Notifications
Number range: 2000 to 2999.
The power sensors report their error states to the R&S NRX. The error states depend
on the power sensor type. Described notifications:
2001 - sensor maximum allowed count
.......................................................................493
2003 - sensor settings conflict
.....................................................................................493
...........................................................................................493
............................................................................................. 494
............................................................................................... 494
2008 - sensor protocol minor mismatch
...................................................................... 494
2009 - sensor protocol major mismatch
...................................................................... 494
2001 - sensor maximum allowed count
The maximum number of power sensors that you can use simultaneously is reached.
Solution: Install the second measurement channel (R&S NRX-K2) or the 3rd and 4th
measurement channel (R&S NRX-K4).
See also:
●
Chapter 5.1, "Parallel Measurements"
●
Chapter 12, "Option Management"
2003 - sensor settings conflict
The current settings of the R&S NRX do not comply with the operating range of the
connected power sensor.
Solution: Change the settings of the R&S NRX or use another power sensor that is
suitable for the selected settings.
2005 - sensor underrange
Possible reasons:
●
The detected signal is below the minimum value that the power sensor can cor-
rectly measure.
●
The current settings of the R&S NRX undercut the specified operating range of the
connected power sensor.