Remote Control Commands
R&S
®
NRX
235
User Manual 1178.5566.02 ─ 08
CALCulate<Measurement>:METer<DirectionalChannel>:LOWer[:DATA]:RATio:
SWR
<value>
Sets the lower limit for the standing wave ratio (SWR) bargraph display.
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
0.0 to 1e18
*RST:
1.0
Default unit: -
Manual operation:
"Forward Scale Lower Limit, Reflection Scale Lower Limit"
CALCulate<Measurement>:METer<DirectionalChannel>:LOWer[:DATA]:RATio[:
VALue]
<value>
Sets the lower limit for the power ratio bargraph display.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
-180.0 to +180.0
*RST:
-20.0
Default unit: dB
Manual operation:
"Forward Scale Lower Limit, Reflection Scale Lower Limit"
CALCulate<Measurement>:METer<DirectionalChannel>:LOWer[:DATA][:POWer]
<value>
Sets the lower limit for the power value bargraph display.
If you enter a value without unit, the unit is defined by
. For further information, see
Measurement Settings and Results