Remote Control Commands
R&S
®
NRX
401
User Manual 1178.5566.02 ─ 08
You can query the defined string using
.
Parameters:
<string>
Identification string. Maximum string length is 128 characters.
Manual operation:
SYSTem:IDN:AUTO
<status>
Enables or disables the automatic instrument identification for
.
Parameters:
<status>
ON | OFF
*RST:
1
Manual operation:
SYSTem:IDN:MODE
<mode>
Sets which identification string is used.
Parameters:
<mode>
AUTO | USER
AUTO
Automatic instrument identification.
USER
Customized identification string. Define the string using
.
*RST:
AUTO
Manual operation:
SYSTem:LANGuage
<language>
Sets the native remote command set or an emulation of a predecessor or a power
meter from another manufacturer. See also
Chapter 14.16, "Remote Emulation"
Setting parameters:
<language>
String
SCPI | NRX
Native remote command set of the R&S NRX.
NRP2 | NRP
Emulation of the selected predecessors.
N432A | N1911A | N1912A | E4418B | E4419B
Requires the NRX KS emulation mode (R&S NRX-K301) option.
Emulation of the selected Keysight power meter.