Remote Control Commands
R&S
®
NRX
273
User Manual 1178.5566.02 ─ 08
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
Alias:
UNIT<Measurement>:POWer:RELative:STATe
CALCulate<Measurement>:RELative<DirectionalChannel>[:MAGNitude]:
AUTO
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<state>
OFF | ON | SET
OFF
Absolute power or power ratio
ON
Relative power or power ratio, using the specified reference
value.
SET
Uses the current measurement value as reference value and
enables the relative measurement.
*RST:
OFF
Manual operation:
"Forward Relative State, Reflection Relative State"
[SENSe<Sensor>:]POWer:REFerence
<ref>
Sets the reference value for the relative power indication.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<ref>
Range:
-290.0 to +110.0
*RST:
+30.0
Default unit: dBm
Manual operation:
"Forward Reference Value, Reflection Reference Value"
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH