Remote Control Commands
R&S
®
NRX
276
User Manual 1178.5566.02 ─ 08
CALCulate<Measurement>:TRACe:X:POINts
................................................................... 276
[SENSe<Sensor>:]BWIDth:VIDeo:LIST?
.......................................................................... 276
[SENSe<Sensor>:]BANDwidth:VIDeo:LIST?
.....................................................................276
[SENSe<Sensor>:]TRACe:REALtime
...............................................................................276
CALCulate<Measurement>:TRACe:X:POINts
<points>
Sets the number of required values per trace sequence.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<points>
Range:
1 to 8192
*RST:
660
[SENSe<Sensor>:]BWIDth:VIDeo:LIST?
[SENSe<Sensor>:]BANDwidth:VIDeo:LIST?
Queries the parameters available for
[SENSe<Sensor>:]BANDwidth:VIDeo
.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
[SENSe<Sensor>:]TRACe:REALtime
<state>
Effective for trace measurements.
If disabled, each measurement from the power sensor is averaged. If enabled, only
one sampling sequence per measurement is recorded, thus increasing the measure-
ment speed. With a higher measurement speed, the measured values of an individual
measurement are immediately delivered.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
ON | OFF
*RST:
OFF
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH