Remote Control Basics
R&S
®
NRX
487
User Manual 1178.5566.02 ─ 08
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Sensor A initializing
Sensor B initializing
Sensor C initializing
Sensor D initializing
NRT Sensor initializing
Other Sensor initializing
0
0
0
0
0
0
0
0
0
+
Figure 15-15: Operation sense status register
Querying the register:
●
STATus:OPERation:SENSe[:SUMMary]:CONDition?
●
STATus:OPERation:SENSe[:SUMMary][:EVENt]?
Table 15-17: Used operation sense status bits and their meaning
Bit no.
Short description
Bit is set if a sensor is initializing.
1
Sensor
A initializing
Sensor
A
2
Sensor
B initializing
Sensor
B
3
Sensor
C initializing
Sensor
C
4
Sensor
D initializing
Sensor
D
5
NRT sensor initializing
R&S NRT-Zxx power sensor
6
Other sensor initializing
USB or LAN power sensor
15.4.8.7
Operation Lower Limit Fail Status Registers
The
CONDition
registers contain information whether a measured value is below a
configured lower limit. The
EVENt
registers contain information whether a measured
value dropped below a limit value since the last query.
For information on index
Chapter 15.4.1, "Hierarchy of the Status Registers"
Status Reporting System