Starting and Ending a Measurement
In a basic scenario, the measurement is started immediately after the measurement
mode is enabled.
If you want to start the measurement only if a specific condition is fulfilled, define a trig-
ger.
Further information:
●
●
Chapter 14.6.2, "Configuring the Trigger"
INITiate<Measurement>:CONTinuous
..............................................................................223
INITiate<Undef>:ALL:CONTinuous
.................................................................................. 223
..................................................................................... 223
..........................................................................................224
INITiate<Measurement>[:IMMediate]
................................................................................224
INITiate<Undef>:ALL[:IMMediate]
.................................................................................... 225
ABORt<Measurement>
Immediately interrupts the current measurement. If the measurement has been started
as a single measurement (
INITiate<Measurement>[:IMMediate]
sensor goes into the idle state. However, if a continuous measurement is in progress
(
INITiate<Measurement>:CONTinuous
ON
), the trigger system of the power sen-
sor enters the waiting for trigger state, and if the trigger condition is met, a new mea-
surement is immediately started.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Event
ABORt<undef>:ALL