
Remote Control Commands
R&S
®
NRX
320
User Manual 1178.5566.02 ─ 08
HIGH
Measured power exceeds the upper scale limit.
*RST:
HIGH
Manual operation:
See
[SENSe<Sensor>:]POWer:REFLection:RANGe:LIMit[:STATe]
<state>
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LIMit[:STATe]
<state>
Enables or disables the Out
1
/
Trig
Out connector as a monitoring output for the power
indication.
If enabled, you cannot use the connector for any other purpose.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
*RST:
OFF
Manual operation:
See
[SENSe<Sensor>:]POWer:REFLection:RANGe:LOWer
<lower>
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LOWer
<lower>
Effective if
[SENSe<Sensor>:]POWer:REFLection:RANGe:AUTO
Sets the lower scale limit for the power indication.
The entry has no unit. The unit corresponds to the output unit selected by
UNIT<Measurement>:POWer[:VALue]
If you change the unit, the entered value remains the same.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<lower>
Range:
-1999.0 to 1999.0
*RST:
0.0
Manual operation:
See
"Measurement for Limit Output"
[SENSe<Sensor>:]POWer:REFLection:RANGe[:UPPer]
<upper>
[SENSe<Sensor>:]POWer[:POWer]:RANGe[:UPPer]
<upper>
Effective if
[SENSe<Sensor>:]POWer:REFLection:RANGe:AUTO
Sets the lower scale limit for the power indication. For further details, see
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LOWer
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH