Remote Control Commands
R&S
®
NRX
246
User Manual 1178.5566.02 ─ 08
Parameters:
<unit>
DBM | DBUV | W
Available units.
If the R&S NRP2 emulation is enabled using
*RST:
DBM
Manual operation:
See
SOURce:UNIT:POWer
<power>
Requires the sensor check source (R&S NRX-B1).
Sets the unit of the power level for the output signal.
Parameters:
<power>
DBM | DBUV | W
*RST:
DBM
Manual operation:
See
UNIT<Measurement>:POWer:RATio
<unit>
Sets the output unit for the measured power ratio values. For NRT measurements, sets
the unit of the relative forward power measurement.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<unit>
DB | DPCT | O
The character
O
stands for One (x1).
*RST:
DB
Manual operation:
See
See
UNIT<Measurement>:POWer:REFLection
<unit>
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:REVerse"
is set.
Sets how the ratio of forward and reverse power is expressed.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<unit>
RCO | RL | SWR | RFR
Measurement Settings and Results