Remote Control Commands
R&S
®
NRX
263
User Manual 1178.5566.02 ─ 08
TRIGger<Measurement>[:CHANnel<Channel>]:JITTer:METHod
<method>
Defines the method how to cope with the misalignment between the trigger event and
the sample point.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<method>
COMPensate | MEASure | NONE
COMPensate
Compensation means resampling of trace result.
MEASure
Does not perform resampling, but stores the measured trigger
jitter.
*RST:
COMPensate
Manual operation:
TRIGger<undef>:ALL:LEVel
<level>
TRIGger<Measurement>[:CHANnel<Channel>]:LEVel
<level>
TRIGger<Measurement>[:CHANnel<Channel>]:SOURce
INTernal
.
Sets the trigger threshold for internal triggering derived from the test signal.
If you enter a value without unit, the unit is defined by
. For further information, see
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<level>
Range:
-290.0 to +223.01
*RST:
-10.0
Default unit: dBm
Manual operation:
Measurement Settings and Results