Remote Control Commands
R&S
®
NRX
264
User Manual 1178.5566.02 ─ 08
TRIGger<undef>:ALL:MODE
<mode>
TRIGger<Measurement>:MODE
<mode>
Controls the trigger execution depending on the setting of the trigger source, see
TRIGger<Measurement>[:CHANnel<Channel>]:SOURce
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<mode>
NORMal | FREerun | SINGle | AUTO
NORMal
Continuous triggering with regular trigger events.
FREerun
Enables a continuous measurement. The power sensor exe-
cutes one measurement cycle after the other.
SINGle
Disables continuous triggering so that only one trigger event at a
time is executed.
AUTO
Automatically starts a measurement if no trigger event has
occurred after 300
ms.
*RST:
AUTO
Manual operation:
See
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer:PORT
<port>
Effective only if the connected sensor is trigger master, see
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer[:STATe]
Sets the port where the trigger master sensor outputs a digital trigger signal.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<port>
INTernal | EXT2 | EXTernal2
*RST:
INTernal
Manual operation:
See
Measurement Settings and Results