Troubleshooting
R&S
®
NRX
494
User Manual 1178.5566.02 ─ 08
Solution: Change the settings of the R&S NRX or use another power sensor that is
suitable for the selected settings.
2006 - sensor overrange
Possible reasons:
●
The detected signal is above the maximum value that the power sensor can cor-
rectly measure.
●
The current settings of the R&S NRX exceed the specified operating range of the
connected power sensor.
Solution: Change the settings of the R&S NRX or use another power sensor that is
suitable for the selected settings.
2007 - sensor overload
The RF input power exceeds the measurement range by far.
Solution: Immediately disconnect the power sensor from the RF source to avoid dam-
age. Use an attenuator or another power sensor that is suitable for the input level.
2008 - sensor protocol minor mismatch
The firmware version of the connected power sensor is out-of-date.
Solution: An update of the power sensor firmware is recommended.
2009 - sensor protocol major mismatch
The firmware version of the connected power sensor is not supported any more.
Solution: Update the firmware of the connected power sensor.
16.2.1.3
Sensor Check Source (R&S NRX-B1) Notifications
Number range: 3000 to 3999. Described notifications:
....................................................................................... 494
3003 - missing calibration data
................................................................................... 494
3004 - version <no.> available for sensor check source (R&S NRX-B1)
3002 - setting not supported
The sensor check source (R&S NRX-B1) does not support the set frequency or power
level.
Solution: Change the frequency or power level setting. See
3003 - missing calibration data
The calibration data of the sensor check source (R&S NRX-B1) are missing for the set
power level.
Solution:
●
Change the frequency or power level setting. See
●
If you cannot work with another power level, servicing is required. Contact cus-
tomer support. See
Chapter 16.5, "Contacting Customer Support"