80 • Start Up, Leak Check Mode
GX-2012 Operator’s Manual
6. If
CL RMNDR
is turned on (factory setting), the screen that appears next depends on
how
CL EXPRD
is set in the Maintenance Mode Menu (page 119).
•
If the unit is due for calibration and
CL EXPRD
is set to CONFIRM (factory
setting), then the following screen displays.
The alarm LED’s and buzzer will pulse several times. After this, press and
release the RESET SILENCE button to continue.
•
If the unit is due for calibration and
CL EXPRD
is set to NOT USE, then the
following screen displays.
The GX-2012 cannot be used until a calibration has been performed either by
selecting
AUTO CAL
, or
ONE CAL
in the Maintenance Mode menu. See
“Calibrating Using Auto Calibration” on page 97 or “Calibrating Using Single
Calibration” on page 100 for calibration instructions.
•
If calibration is not due or
CL EXPRD
is set to NO EFFECT, then the following
screen appears for a few seconds indicating when the next calibration is due.
CAL
C--LIMIT
FAIL
C--LIMIT
2011
9
30
69d
NEXT CAL