GX-2012 Operator’s Manual
Using Maintenance Mode • 107
•
To return to the main menu, press POWER ENTER at any time. If any channel
failed the calibration, the buzzer will sound a double pulsing tone until you press
and release the POWER ENTER button.
•
Disconnect the tubing from the probe.
13. Unscrew the demand flow regulator from the calibration cylinder.
14. With the
BUMP
menu option displayed, press the (SHIFT)
button until the
NORMAL
menu option is displayed.
15. Press and release the POWER ENTER button to return to Measuring Mode.
Turning the Lunch Break Setting On or Off
With
LNCH BRK
set to
OFF
(factory setting), the GX-2012 automatically starts new
TWA and PEAK reading collection and resets the time in operation at startup.
With
LNCH BRK
set to
ON
, the Resume Measurements Screen displays during startup.
From this screen, you can choose to continue accumulating TWA and PEAK readings and
the time in operation from the last time the GX-2012 was used or start collecting new
readings and reset the time in operation.
1. Use the AIR
or (SHIFT)
buttons to display the
LNCH BRK
menu item.
49
12.0
5
25.0
%LEL
%
ppm
ppm
MAINTENANCE
CH
4
O
2
/CAL
CO
H
2
S
MAINTENANCE
BUMP
MAINTENANCE
LNCH BRK