138 • Replacing or Recharging the Batteries
GX-2012 Operator’s Manual
Figure 23: Putting the GX-2012 into the Charging Station
5. When a full charge has been reached, approximately 4 hours, the red LED on the
charging station will turn off. Remove the GX-2012 from the charging station and
unplug the charging station’s power cord from the AC outlet.
Recharging the Lithium Ion Battery Pack Out of the Instrument
The lithium ion battery pack may be charged using the charging station while it is out of
the GX-2012. This is useful if spare battery packs are kept in case the pack in the GX-
2012 needs to be charged, but the unit must be used immediately. In this case, a spare
charged pack can be installed in the GX-2012 and the dead pack charged in the charging
station.
1. Make sure the GX-2012 is off.
2. Confirm that the adapter’s DC output cord is plugged into the charger. If a multi-
instrument charging station is being used, the same AC adapter is used as for the
single instrument charging station.