128 • Using Maintenance Mode
GX-2012 Operator’s Manual
ROM/SUM
The
ROM/SUM
screen shows the firmware version that is loaded in the instrument and
the firmware checksum.
1. Use the AIR
button to scroll to the ROM/SUM screen. Press POWER ENTER and
the ROM and SUM values for your unit will be displayed.
The first two lines display the ROM number. The ROM number indicates the
firmware version number. In the above example, the ROM number is 51513. The
bottom line displays the SUM number, 7864 in the above example.
2. Press POWER ENTER again to return to the ROM/SUM menu screen.
Turning the Password Function On or Off
With
PASSWORD
set to
ON
, the GX-2012 prompts you for a password when you enter
Calibration Mode.
With
PASSWORD
set to
OFF
(factory setting), no password is required to enter
Calibration Mode.
NOTE:
A password is always required to enter Maintenance Mode even if
PASSWORD
is turned OFF. You may change the factory set password by
turning
PASSWORD
ON and entering a new password when prompted for it.
1. Use the AIR
or (SHIFT)
buttons to display the
PASSWORD
menu item. Press
and release the POWER ENTER button. The screen indicates the current setting.
2. Use the AIR
or (SHIFT)
button to display the desired setting.
5
1513
7864
MAINTENANCE
ROM/SUM
On
MAINTENANCE
PASSWORD
www.calcert.com