GX-2012 Operator’s Manual
Calibration Mode • 61
17% oxygen can be used to make the adjustment.
NOTE:
If your combustible gas channel is calibrated to a gas other than
methane, use an appropriate gas cylinder to perform the calibration.
CAUTION:
The single calibration method does not have a “FAIL” notification. If a
sensor cannot be set to agree with the calibration source, see
“Troubleshooting” on page 131.
1. While in the calibration menu, use the AIR
or (SHIFT)
button to navigate to the
ONE CAL menu option.
2. Press and release the POWER ENTER button to display the single calibration menu.
The %LEL sensor will be displayed first.
3. Use the AIR
or (SHIFT)
button to display the sensor you want to calibrate (in
this example the combustible gas % LEL sensor). If you have a 5 sensor unit, the
%vol sensor will be one of the choices. It is shown below.
MAINTENANCE
ONE CAL
- - -
%LEL
MAINTENANCE
CH
4
ONE CAL
- - -
vol%
MAINTENANCE
CH
4
ONE CAL