156 • Calibrating the GX-2012
GX-2012 Operator’s Manual
7. Press and release the POWER ENTER button to save the span value. The LCD will
indicate that the calibration has ended, then the single calibration menu displays.
8. Disconnect the tubing from the GX-2012’s probe.
9. Repeat steps 3 through 8 for any other sensors you want to calibrate. Make sure you
use an appropriate calibration cylinder for each sensor.
CAUTION:
When calibrating the oxygen channel, verify the concentration of oxygen
listed on the cylinder’s label. For oxygen-free samples (100% nitrogen for
example), set the oxygen zero setting to 0.0%.
10. After the last channel is calibrated, disconnect the calibration tubing from the probe.
11. Use the AIR
and (SHIFT)
buttons to scroll to ESCAPE in the ONE CAL menu.
12. Press and release the POWER ENTER button to exit ONE CAL.
13. With the ONE CAL menu option displayed, press the (SHIFT)
button until the
NORMAL
menu option is displayed.
14. Press and release the POWER ENTER button to return to Measuring Mode.