122 • Using Maintenance Mode
GX-2012 Operator’s Manual
1. Use the AIR
or (SHIFT)
buttons to display the
BP EXPRD
menu item. Press
and release the POWER ENTER button. The screen will indicate the current setting.
2. Use the AIR
or (SHIFT)
button to display the desired setting along the bottom of
the screen.
3. Press and release the POWER ENTER button to save the setting and return to the
main menu.
Updating the Bump Test Check Setting
This item setting defines which sensors are used to update the bump test reminder screen.
B--CHECK
can be set to 4-GAS or ALL. If
B--CHECK
is set to 4-GAS (factory setting),
the calibration dates for only the standard 4 sensors (LEL, O
2
, H
2
S, and CO) are used to
determine if a bump test is due. If
B--CHECK
is set to ALL, the calibration dates for the
standard 4 sensors along with the %volume sensor are used to determine if a bump test is
due.
1. Use the AIR
or (SHIFT)
buttons to display the
B--CHECK
menu item. Press
and release the POWER ENTER button. The screen will indicate the current setting.
2. Use the AIR
or (SHIFT)
button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting and return to the
main menu.
tESt
SEL
MAINTENANCE
CONFIRM
4GAS
MAINTENANCE
B--CHECK