112 • Using Maintenance Mode
GX-2012 Operator’s Manual
7. Press POWER ENTER to save the setting. The unit will direct you back to the Alarm
Points Screen where you can use the AIR
or (SHIFT)
buttons to select the
oxygen channel.
Press POWER ENTER.
8. The first selection for the oxygen channel is the alarm pattern.
The choices are L-H, L-LL, or H-HH. The L-H pattern provides a decreasing warning
and an increasing alarm. The L-LL pattern provides a decreasing warning and a
decreasing alarm. The H-HH pattern provides an increasing warning and an
increasing alarm. This setting is for operation in inert atmospheres and will cause
INERT to appear along the bottom line of the display during Normal Operation. Make
your selection using the AIR
or (SHIFT)
buttons and press POWER ENTER
when finished. The factory setting is L-H.
9. You may then adjust the warning setting and alarm setting for the oxygen channel
using the AIR
or (SHIFT)
buttons. Press POWER ENTER when finished with
each selection to return to the Alarm Points Screen.
10. Use the AIR
or (SHIFT)
buttons to select the CO channel and press POWER
ENTER to display the warning setting.
11. Adjust the values for the warning and alarm settings using the procedure in step 4
through step 7. When you press POWER ENTER after selecting the alarm value, you
will then be prompted to select the STEL value.
- - -
%
MAINTENANCE
O
2
ALARM--P
L-H
MAINTENANCE
O
2
ALM PTN
200
ppm
MAINTENANCE
STEL
CO