GX-2012 Operator’s Manual
Calibration Mode • 67
If any channel fails the bump test,
•
The instrument will beep and a calibration is immediately and automatically
started. Continue to apply the calibration gas. CAL on the bottom left of the
screen will alternate with APPLY.
The calibration time will begin to countdown in seconds in the lower right of the
screen. The calibration time counted down during a calibration initiated because
of a failed bump test is the difference between the bump test gas time and the
calibration time defined in the
BUMP--SET
menu item in Maintenance Mode.
•
At the end of the calibration, the instrument displays the results from both the
bump test and the calibration. Use the AIR
or (SHIFT)
button to scroll
between the calibration/bump test results, the bump test gas readings, and the
calibration gas readings.