GX-2012 Operator’s Manual
Start Up, Bar Hole Mode • 71
•
If calibration is not due or
CL EXPRD
is set to NO EFFECT, then the following
screen appears for a few seconds indicating when the next calibration is due.
7. If
BP RMNDR
is turned on (factory setting is OFF), the screen that appears next
depends on how
BP EXPRD
is set in the Maintenance Mode Menu (page 119).
•
If the unit is due for a bump test and
BP EXPRD
is set to CONFIRM (factory
setting), then the following screen displays.
The alarm LED’s and buzzer will pulse several times. After this, press the
RESET SILENCE button until you hear a beep to continue.
•
If the unit is due for a bump test and
BP EXPRD
is set to NOT USE, then the
following screen displays.
The GX-2012 cannot be used until a bump test has been performed by selecting
BUMP
in the Maintenance Mode menu. See “Performing a Bump Test” on
page 102 for bump testing instructions.