GX-2012 Operator’s Manual
Using Maintenance Mode • 103
6. Press and release the POWER ENTER button to display the bump test gas values.
7. Use the AIR
or (SHIFT)
buttons to see the %volume combustible gas bump test
value.
NOTE:
If you wish to change any of the bump test gas values, you must exit the
BUMP
menu item and enter the
AUTO CAL
menu item. See steps 8 through 15 in
“Calibrating Using Auto Calibration” on page 97 for instructions to change the
gas values.
8. A bump test may be performed on the standard 4 sensors or on the %volume
combustible sensor. Use the AIR
or (SHIFT)
buttons to display the screen for the
sensor(s) you wish to bump test.
9. Before proceeding, confirm that the bump test gas value(s) are the same as the
concentration(s) in the calibration cylinder. If they are not, adjust the bump test gas
value(s) by entering the AUTO CAL menu item, changing the values there, and
reentering the BUMP menu.
10. Connect the tubing from the demand flow regulator to the rigid tube on the probe and
quickly press and release the POWER ENTER button. The unit will begin counting
down from 30.
50
12.0
50
25.0
%LEL
%
ppm
ppm
MAINTENANCE
CH
4
O
2
BUMP 30
CO
H
2
S
100
vol%
MAINTENANCE
CH
4
BUMP 30