4. Installation
4-7 Connecting the power and signal cables
63 / 168
(The numbers are
terminal numbers.)
5 Connect the power and signal cables to the terminal plate.
Connect the cables to the following terminals on the
terminal plate. For information on how to connect cables to
the terminal plate, refer to ‘4-7-5 Connecting cables to the
terminal plate’.
Terminal
No.
Power/signal cable connection
1
Power supply (+)
2
Power supply (−)
3
Signal (+)
4
Signal (−)
Recommended insulation length: 40 - 50 mm
When connecting cables, connect the cable shielding to
the terminals at the top left or top right on the inside of the
main unit.
Depending on the installation location, noise interference
may result.
In such cases, remove the shielding.
6 Route the cable from inside the main unit to the
connector.
Route to the connector on the side where the cable gland
or conduit is attached.
7 Mount the terminal plate on the terminal plate board.
Adjust the cable position so that no load is applied to the
terminal plates.
<With cable (external cable) routed to right-hand side>
<With cable (external cable) routed to left-hand side>
Sheathing
Conductor
Insulation
1
2
3
4
Cable gland
Содержание GD-3EC
Страница 28: ...2 Important Safety Information 2 4 Safety information 28 168 Connection examples GD 3RI GD 3ECB SD 3SC SD 3SC...
Страница 169: ...11 Appendix Revision History 169 168...
Страница 170: ...11 Appendix Revision History 170 168...
Страница 171: ...11 Appendix Revision History 171 168...
Страница 172: ...11 Appendix Revision History 172 168...
Страница 173: ...11 Appendix Revision History 173 168...
Страница 174: ...11 Appendix Revision History 174 168...
Страница 175: ...11 Appendix Revision History 175 168...
Страница 176: ...11 Appendix Revision History 176 168...
Страница 177: ...11 Appendix Revision History 177 168...