9. Troubleshooting
9-2 Reading abnormalities
139 / 168
Symptom/display
FAULT
lamp
Cause
Action
Sudden environmental
changes
Alarm indications may be issued by the product
due to sudden environmental fluctuations (e.g.,
temperature).
The product cannot be used in locations where
sudden environmental changes occur frequently.
The user should take appropriate measures.
The sensor sensitivity
has degraded.
Replace with a new sensor.
The calibration gas
concentration is
inappropriate.
Prepare the correct calibration gas.
The sensor sensitivity
has degraded.
Replace with a new sensor.
System
abnormality
(E-9)
―
Clock abnormality
Set the date and time.
Correction may not function correctly if the EC
sensitivity correction function is enabled.
If similar symptoms recur frequently, the internal
clock may be defective. Replace with a new
internal clock.
Contact Riken Keiki.
Outside operating
temperature
range
(E-27)
―
Ambient temperatures at
the location where the
product is installed are
outside the operating
temperature range.
The product operating temperature range varies
with the specific detection principle. The product
cannot be used outside the operating
temperature range.
The user should take appropriate measures.
Sensor life
expiration
warning
(E-8)
―
The sensor is
approaching the end of
its service life.
We recommend sensor replacement.
Contact Riken Keiki.
Sensor warning
(E-1)
―
Contact Riken Keiki.
Содержание GD-3EC
Страница 28: ...2 Important Safety Information 2 4 Safety information 28 168 Connection examples GD 3RI GD 3ECB SD 3SC SD 3SC...
Страница 169: ...11 Appendix Revision History 169 168...
Страница 170: ...11 Appendix Revision History 170 168...
Страница 171: ...11 Appendix Revision History 171 168...
Страница 172: ...11 Appendix Revision History 172 168...
Страница 173: ...11 Appendix Revision History 173 168...
Страница 174: ...11 Appendix Revision History 174 168...
Страница 175: ...11 Appendix Revision History 175 168...
Страница 176: ...11 Appendix Revision History 176 168...
Страница 177: ...11 Appendix Revision History 177 168...