8. Storage, Relocation, and Disposal
8-1 Procedures for storage or when not in use for extended periods
135 / 168
<Storage conditions (Sensor-fitted SD-3)>
Model
Storage
limit
(years)
Temperature
(°C)
Humidity
*1
(% RH)
Pressure
(kPa)
Remarks
-
Minimum
Maximum
Minimum
Maximum Minimum Maximum
SGF
0.5
-10
40
0
90
80
120
・
Store in a location away from direct
sunlight.
・
Store in an environment free of
corrosive gas, vibration, and dust.
・
Avoid condensation and moisture.
・
Store in an environment free of
gases that may act as a catalyst
poison (e.g., silicone, chlorine, sulfur
gases).
SHF
0.5
-10
40
0
90
80
120
NCF
0.5
-10
40
0
90
80
120
IRF
0.5
-10
40
0
90
80
120
・
Store in a location away from direct
sunlight.
・
Store in an environment free of
corrosive gas, vibration, and dust.
・
Avoid condensation and moisture.
ESF-A
*2
0.5
-10
40
0
90
80
120
・
Store in a location away from direct
sunlight.
・
Store in an environment free of
corrosive gas, vibration, and dust.
・
Avoid condensation and moisture.
・
Store in an environment free of
gases that may act as a catalyst
poison (e.g., silicone, chlorine, sulfur
gases).
・
Store in the dedicated sensor storage
container.
*5
Open the dedicated container just
before use, remove the sensor, and
mount in the detector.
ESF-X
*3
ESF-B
*4
0.5
-10
40
0
90
80
120
*1: Maximum humidity corresponds to an absolute humidity of 33 g/m3 (33.6 °C, 90 %RH).
*2: ESF-A refers to sensor model names starting with “ESF-A”.
*3: ESF-X refers to sensor model names starting with “ESF-X”.
*4: ESF-B refers to sensor model names starting with “ESF-B”.
*5: If not stored in the container, fluid leakage may occur under high humidity conditions of 90 %RH or more. Fluid
shortage may occur with the ESF-B
*4
under low humidity conditions.
CAUTION
This product is very heavy and features various projections. Damage to this product and accompanying
parts may result if they are packaged in general packaging materials. When storing or returning the
product to Riken Keiki for any reason, use the packaging in which the product was shipped. If the original
packaging is no longer available, wrap twice and place hard objects in the spaces within the packaging to
ensure that the product will not shift inside the packaging. Avoid packaging fragile items.
Содержание GD-3EC
Страница 28: ...2 Important Safety Information 2 4 Safety information 28 168 Connection examples GD 3RI GD 3ECB SD 3SC SD 3SC...
Страница 169: ...11 Appendix Revision History 169 168...
Страница 170: ...11 Appendix Revision History 170 168...
Страница 171: ...11 Appendix Revision History 171 168...
Страница 172: ...11 Appendix Revision History 172 168...
Страница 173: ...11 Appendix Revision History 173 168...
Страница 174: ...11 Appendix Revision History 174 168...
Страница 175: ...11 Appendix Revision History 175 168...
Страница 176: ...11 Appendix Revision History 176 168...
Страница 177: ...11 Appendix Revision History 177 168...