5. Usage Instructions
5-2 Startup
81 / 168
CAUTION
Never turn off the power while initialization is underway. The internal memory is read during initialization.
A specific warmup is required for each sensor type at startup and after sensor replacement. Allow the
product to warm up for the specified time. Expected warmup completion times for the semiconductor type
(SGF) and hot-wire semiconductor type (SHF) sensors are indicated after initialization. Alarm activation
and output signals will be unstable during warmup. Notify the relevant department beforehand to avoid
conditions that may lead to errors.
Perform calibration once warmup is complete.